Strong cleaner for the preparation of instruments in ultrasonic devices


  • Strong cleaning power
  • Active disinfectant
  • Effective against Aspergillus niger spores
ULTRACLEAN-3 is a concentrate for the cleaning of surgical and dental instruments (e.g. scalpels, tooth extracting forceps, impression trays, clamps, scissors, micro instruments, spatulas, bone files, probes, excavators, scalers) in the ultrasonic cleaning bath. The alkaline-based concentrate successfully removes residual protein, blood, even cement and plaster. ULTRACLEAN-3 contains highly efficient disinfecting agents active against bacteria and fungi including the extremely resistant spores of Aspergillus niger.

Use

For instrument cleaning, use an aqueous solution and dosage as required. To measure the correct dosage, use lid as a measuring cup. Add liquid concentrate to the cup until liquid reaches desired scale ml unit (20 ml lower scale, 40 ml upper scale). Use an ultrasound cycle of 30 minutes and then rinse instruments thoroughly with water. ULTRACLEAN-3 is not suitable for the cleaning of color-coded instruments and instruments fully/partly made of/with natural bristles, nylon or aluminum.

Scope & Application

  • For the cleaning of surgical and dental instruments in the ultrasonic cleaning bath
  • Scalpels, forceps, spoons, clamps, scalars, micro-instruments, spatulas, files, curettes, and probes
  • Not suitable for the cleaning of color-coded instruments and instruments fully / partly made of / with natural bristles, nylon or aluminum

Effective spectrum / Exposure times

Bactericide5% in 15 min.
Fungicide5% in 30 min.

Complies with European Standards EN 1040, EN 1275

Dosing

5%
100 ml
1900 ml
30 min.

Material Compatibility

 Polyethylenecompatible
 PTFEcompatible
 Polypropylenecompatible
 Cast-ironcompatible

Frequently Asked Questions

  • Can ULTRACLEAN-3 be used as an instrument disinfectant?
    No. Even though ULTRACLEAN-3 has disinfecting properties and works against bacteria and fungi including the extremely resistant spores of Aspergillus Niger, it is only designed to be outstanding instrument cleaner.
  • Must I rinse instruments after cleaning them with ULTRACLEAN-3?
    Yes. Rinsing with water after disinfection with ULTRACLEAN-3 is required.
  • How frequently should I change the solution?
    Depending on the organic contamination, change the solution 2-3 times per week. However, a daily change is recommended.

Composition in 100 g

16 g Potassium hydroxide, Complexing agent, Special surfactants.

Physical properties

Color: Light-yellow liquid
Odour: Aromatic
Shelf life: 60 months
pH-Value 5% solution: 13.0

Precautions

R35 Causes severe burns. S2 Keep out of the reach of children. S26 In case of contact with eyes, rinse immediately with plenty of water and seek medical advice. S28 After contact with skin, wash immediately with plenty of water. S37/39 Wear suitable gloves and eye/face protection. S45 In case of accident or if you feel unwell, seek medical advice immediately (show the label where possible).

Delivery Forms

Article-Nr. Product
CH-061050 ULTRACLEAN-3 1 liter